Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-229
smuX.trigger.autoclear
This attribute turns automatic clearing of the event detectors on or off.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Not saved
smuX.DISABLE
Usage
autoClear
= smu
X
.trigger.autoclear
smu
X
.trigger.autoclear =
autoClear
autoClear
Auto clear setting. Set to one of the following values:
0
or
smu
X
.DISABLE
: Turns off automatic clearing of the event detectors
1
or
smu
X
.ENABLE
: Turns on automatic clearing of the event detectors.
X
SMU channel (for example,
smua.trigger.autoclear
applies to SMU Channel
A).
Details
This attribute is used to enable or disable automatic clearing of the trigger model state machine event detectors
when the SMU transitions from the arm layer to the trigger layer.
Only the detected state of the event detectors will be cleared.
The overrun status of the event detectors are not automatically cleared when the SMU transitions from the arm
layer to the trigger layer.
The event detectors are always cleared when a sweep is initiated.
Also see