![Keithley 2651A Reference Manual Download Page 565](http://html1.mh-extra.com/html/keithley/2651a/2651a_reference-manual_661729565.webp)
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-257
measurementRegister
The measurement event instrument summary register status. A zero (0)
indicates no bits set (also send 0 to clear all bits); other values indicate various
bit settings
Details
These attributes are used to read or write to the measurement event instrument summary registers. Reading a
status register returns a value. The binary equivalent of the returned value indicates which register bits are set.
The least significant bit of the binary number is Bit B0, and the most significant bit is Bit B15. For example,
assume the value 2 is returned for the enable register. The binary equivalent is 0000 0000 0000 0010. This value
indicates that bit B1 (SMUA) is set.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
(on page E-
Enable and transition registers
(on page E-18). The individual bits of this register have the following
meanings:
Bit Value
Description
B0
Not used
Not applicable.
B1
status.measurement.instrument.SMUA
Set bit indicates one or more enabled bits of the
measurement event SMU A summary register
is set.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2-B15
Not used
Not applicable.
As an example, to set bit B1 of the measurement event instrument summary enable register, set
status.measurement.instrument.enable = status.measurement.instrument.SMUA
.
In addition to the above constant,
measurementRegister
can be set to the decimal equivalent of the bit to set.
Example
status.measurement.instrument.enable =
status.measurement.instrument.SMUA
Sets the SMU A bit of the
measurement event instrument
summary enable register using a
constant.
Also see
(on page E-23)
status.measurement.instrument.smuX.*
This attribute contains the registers of the measurement event SMU X summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
387 (All bits set)