![Keithley 2651A Reference Manual Download Page 299](http://html1.mh-extra.com/html/keithley/2651a/2651a_reference-manual_661729299.webp)
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 6: Instrument programming
2651A-901-01 Rev. A / March 2011
6-55
The second method to run parallel test scripts is to use groups with a single TSP-Link network. A
group consists of one or more nodes with the same group number. Each group on the TSP-Link
network can run different test scripts at the same time (in parallel).
The following figure displays a single TSP-Link network with groups. This method requires one
TSP-Link network and a single GPIB connection to the computer.
Figure 91: Model 2651A single TSP-Link network with groups
GPIB
TSP-Link In
TSP-Link Out
GPIB
TSP-Link In
TSP-Link Out
GPIB
TSP-Link In
TSP-Link Out
•
•
•
GPIB
TSP-Link In
TSP-Link Out
GPIB
TSP-Link In
TSP-Link Out
GPIB
TSP-Link In
TSP-Link Out
SMU B
GPIB
TSP-Link In
TSP-Link Out
SMU A
GPIB
TSP-Link In
TSP-Link Out
SMU C
GPIB
TSP-Link In
TSP-Link Out
SMU A
GPIB
TSP-Link In
TSP-Link Out
SMU
n
SMU
n
= last SMU in system.
Node
n
= last node in system.
GPIB
TSP-Link In
TSP-Link Out
SMU D
GPIB
Master
Node 3
Node 4
Node 5
Node
n
Node 2
Group leader
Node 6
Group leader
TSP-Link In
TSP-Link Out
SMU A
To PC
Group 1
Group 2
Group 0
The following table describes the functions of a single TSP-Link network. Each group in this example
runs multiple test scripts at the same time (parallel processing).