Source-measure setup
SRC
Selects the source function (V or A) and places the cursor in the source field for editing.
MEAS
Cycles through measure functions (V, A,
Ω,
or W).
LIMIT
Places the cursor in the compliance limit field for editing. Also selects the limit value to edit (V, A, or
W).
MODE
Selects a meter mode (I-METER, V-METER, OHM-METER, or WATT-METER).
Performance control
DIGITS
Cycles through display resolution (4-1/2, 5-1/2, or 6-1/2 digits).
SPEED
Sets the measurement speed (FAST, MEDium, NORMAL, HI-ACCURACY, or OTHER). Setting the
speed selects either the fast or integrating A/D converter. When FAST is selected, the fast A/D
converter is used. When the integrating A/D converter is selected, the measurement speed and
accuracy are set by controlling the measurement aperture. Also see
(on page 2-71).
REL
Controls relative measurements, which allows a baseline value to be subtracted from a reading.
FILTER
Enables or disables the digital filter. You can use this filter to reduce reading noise.
Special operation
LOAD
Loads test for execution (FACTORY, USER, or SCRIPTS).
RUN
Runs the last selected factory or user-defined tests.
STORE
Accesses reading buffers and takes readings. TAKE_READINGS: Sets number of readings to make
and also takes and stores readings. SAVE: Stores readings in one of two internal buffers
(SMUA_BUFFER1 or SMUA_BUFFER2), or to a user-installed flash drive (USB1) in CSV or XML
format. Readings include measurements, source values, and time stamp values, if configured.
RECALL
Recalls information (DATA or STATISTICS) stored in either CHANA-BUFF1 or CHANA-BUFF2:
DATA includes stored readings, and if configured, source values and timestamp values;
STATISTICS includes MEAN, STD DEV, SAMPLE SIZE, MINIMUM, MAXIMUM, PK-PK.
TRIG
Triggers readings.
MENU
Accesses the
(on page 2-16). The main menu can be used to configure many facets of
operation.
EXIT
Cancels selection and backs out of the menu structure. Also used as a
LOCAL
key to take the unit
out of remote operation.
ENTER
Accepts selection and moves to the next choice or exits the menu.
Numbers
Number keys
When enabled, the number keys (0-9, +/-, 0000) allow direct numeric entry in the EDIT mode. Press
the navigation wheel
to enter EDIT mode. For more information, see
(on page 2-