![Keithley 2651A Reference Manual Download Page 598](http://html1.mh-extra.com/html/keithley/2651a/2651a_reference-manual_661729598.webp)
Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-290
2651A-901-01 Rev. A / March 2011
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
(on page E-
Enable and transition registers
(on page E-18). The individual bits of this register have the following
meanings:
Bit Value
Description
B0-B9 Not
used
Not applicable
B10
status.operation.instrument.tsplink.TRIGGER_OVERRUN
status.operation.instrument.tsplink.TRGOVR
Set bit indicates one or
more enabled bits for the
operation status TSP-
Link overrun register is
set.
Bit B10 decimal value:
1024
Binary value:
0100 0000 0000
B11-
B15
Not used
Not applicable
In addition to the above constants,
operationRegister
can be set to the numeric equivalent of the bit to set.
For example, to set bit B10, set
operationRegister
to 1024.
Example
status.operation.instrument.tsplink.enable = 1024
Sets the trigger overrun bit of
the operation status TSP-Link
summary enable register
using a decimal value.
Also see
status.operation.instrument.tsplink.trigger_overrun.*
status.operation.instrument.tsplink.trigger_overrun.*
This attribute contains the operation status TSP-Link overrun register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
14 (All bits set)