Section 2: General operation
Model 2651A High Power System SourceMeter® Instrument Reference Manual
2-38
2651A-901-01 Rev. A / March 2011
WARNING
Hazardous voltages may be present on the output and guard terminals. To prevent electrical
shock that could cause injury or death, NEVER make or break connections to the Model 2651A
while the output is on. Power off the equipment from the front panel or disconnect the main
power cord from the rear of the Model 2651A before handling cables connected to the outputs.
Putting the equipment into standby does not guarantee the outputs are not powered if a
hardware or software fault occurs.
Contact check commands
The following table summarizes basic contact check commands. See
(on page
5-1) for more information on using these commands.
Basic contact check commands
Command Description
flag = smua.contact.check()
Determine if contact resistance is lower than
threshold.
rhi, rlo = smua.contact.r()
Measure the contact resistance.
smua.contact.speed = speed_opt
Set speed_opt to one of the following:
0
or
smua.CONTACT_FAST
1
or
smua.CONTACT_MEDIUM
2
or
smua.CONTACT_SLOW
smua.contact.threshold = rvalue
Resistance threshold for the contact check
function.
Figure 10: Contact check measurement connections
G
G G G
S
LO
SENSE/GUARD
S
HI
LO
HI
OUTPUT
40V, 50A MAX.
S LO
LO
HI
S HI
DUT
R
S
R
C
R
S
R
C
R
S
R
C
Contact
Resistance
Cable/Relay
Resistance