The safety shield can be metallic or nonmetallic, and must completely surround the DUT test circuit. A
metal safety shield must be connected to a known safety earth ground and chassis ground. See
(on page 2-57) later in this section for important safety information on the use of a metal or
nonmetallic enclosure.
Model 2651A safety shield
The maximum output voltage for a Model 2651A channel is 40 V, which is considered a
nonhazardous level. However, using two Model 2651A voltage sources in a series configuration or
(on page 2-58) can cause test circuit voltage to exceed 42 V. In the following figure,
the source-measure units (SMUs) of two Model 2651A instruments can be connected in series to
apply 80 V to a device under test (DUT) (see
(on page 6-53) for information
on using multiple Model 2651A).
Use #18 AWG wire or larger for connections to safety earth ground and chassis.
Figure 21: Safety shield for hazardous voltage using two Model 2651A instruments