Output-off function
This setting is used only when the when output is turned off and the Model 2651A is in NORMAL
output-off mode (
smu
X
.source.offmode = smu
X
.OUTPUT_NORMAL
).
When the Model 2651A is in NORMAL output-off mode, you can set the output-off function to
CURRENT or VOLTAGE through the CONFIG menu on the front panel, or by using the
smu
X
.source.offfunc
attribute from a remote interface. VOLTAGE is the default output-off
function.
When the output is turned off and the selected output-off function is VOLTAGE
(
smua.source.offfunc = smua.OUTPUT_DCVOLTS
):
•
The source-measure unit (SMU) sources 0 V.
•
The current limit is set by the
smua.source.offlimiti
attribute (default 1 mA).
When the output is turned off and the selected output-off function is CURRENT
(
smua.source.offfunc = smua.OUTPUT_DCAMPS
):
•
The SMU sources 0 A.
•
The voltage limit is set by the
smua.source.offlimitv
attribute (default 40 V).
When the output-off function is set to either voltage or current, the SMU may source or sink a very
small amount of power. In most cases, this source or sink power level is not significant.
Selecting the output-off function
NOTE
This setting is used only when the when output is turned off and the source-measure unit (SMU) is in
NORMAL output-off mode.
Setting the output-off function from the front panel
To configure the output-off function from the front panel:
1. Press
the
CONFIG
key.
2. Press
the
OUTPUT ON/OFF
control. The CONFIGURE OUTPUT A menu displays.
3. In the CONFIGURE OUTPUT A menu, select
OFF-STATE
to display the OUTPUT OFF STATE
A menu.
4. With the OUTPUT OFF STATE A menu displayed, select
FUNCTION
to display the OFF
FUNCTION A menu.
5. In the OFF FUNCTION A menu, select
CURRENT
or
VOLTAGE
.