9-6
Digital I/O Port, Interlocks, and Pulse Sync Output
Model 2520 User’s Manual
Interlock operation
shows typical connections to the REMOTE INTERLOCK connector. If the fix-
A), the three source outputs are enabled and can be turned
on. If the lid of the test fixture opens (
B), the switch opens, and the three source
outputs are disabled and cannot be turned on.
Figure 9-4
Interlock operation
Pin 9
Interlock Switch
Closed
Model 2520 Testhead
Test Fixture
A. Model 2520 outputs can be turned on.
Interlock Switch
Open
Test Fixture
B. Model 2520 outputs cannot be turned on.
Remote
Interlock
Pin 1
Model 2520 Testhead
Pin 1
Pin 9
Remote
Interlock
Summary of Contents for 2520