14-50
SCPI Command Reference
Model 2520 User’s Manual
Read condition registers
CONDition?
:STATus:MEASurement:CONDition?
Read Measurement Condition
:STATus:QUEStionable:CONDition?
Read Questionable Register
:STATus:OPERation:CONDition?
Read Operation Condition
Description
These query commands are used to read the contents of the condition
registers.
Select default conditions
PRESet
:STATus:PRESet
Return registers to default conditions
Description
When this command is sent, the following SCPI event registers are
cleared to zero (0):
1. Operation Event Enable Register.
2. Event Enable Register.
3. Measurement Event Enable Register.
NOTE
The Standard Event Register is not affected by this command.
Error queue
[:NEXT]?
:STATus:QUEue[:NEXT]?
Read Error Queue
Description
As error and status messages occur, they are placed into the Error
Queue. This query command is used to read those messages. See
NOTE
The :STATus:QUEue[:NEXT]? query command performs the same function as
the :SYSTem:ERRor? query command. See
“SYSTem subsystem,” page 14-51
CLEar
:STATus:QUEue:CLEar
Clear Error Queue
Description
This action command is used to clear the Error Queue of messages.