2-10
Connections
Model 2520 User’s Manual
Connection considerations
When making connections to the laser diode, observe the following considerations to
avoid pulse degradation due to distributed inductance and other effects:
•
Use only the supplied 15
Ω
coaxial cables.
•
Keep cable lengths to a minimum.
•
Connect the four cable shields together at the DUT.
•
Carry cable shields as close to the DUT as possible. Minimize the length of
exposed (unshielded) signal lines.
•
Dress the VOLTAGE SENSE cables as far away from the CURRENT OUTPUT
cables to avoid magnetic coupling. Twist the SENSE cables together to further
reduce magnetic coupling.
•
Connect the VOLTAGE SENSE leads as close to the body of the DUT as possible.
NOTE
See
, “
,” for information on these and
other possible measurement problems and how to avoid them.