Model 2520 User’s Manual
Digital I/O Port, Interlocks, and Pulse Sync Output
9-3
Start-of-test (SOT) line
The input line (SOT) is a TTL-compatible logic line used by external equipment to start a
test. With the
↓
STEST trigger event selected, a low SOT pulse starts the testing process.
With the
↑
STEST trigger event selected, a high SOT pulse starts the testing process. With
the
↑↓
STEST trigger event selected, a high or low SOT pulse starts the testing process.
See Section 8 for triggering information,
NOTE
SOT input pulses are referenced to ground (pin 9).
+5V output
The Digital I/O Port provides a +5V output that can be used to drive external logic cir-
cuitry. Maximum current output for this line is 300mA. This line is protected by a self-
resetting fuse.
Digital output configuration
There are two basic methods to connect external components to the digital output lines,
sink operation and source operation.
Sink operation
shows the basic output configuration for sink operation. Note that the external
relay coil is connected between the digital output line (pins 1 to 4) and +5V (pin 7). With
this configuration, the digital output line must be set LO to energize the relay, and the
maximum sink current is 500mA.
Figure 9-2
Sink operation
Model 2520
External
Relay
To Other
Circuits
+5V
Maximum Sink Current: 500mA
Pin 7
Pins 1-4
Pin 9
Digital I/O
Port
Summary of Contents for 2520
Page 250: ...A Specifications ...
Page 254: ...B StatusandErrorMessages ...
Page 262: ...C DataFlow ...
Page 266: ...D IEEE 488BusOverview ...
Page 281: ...E IEEE 488andSCPI ConformanceInformation ...
Page 285: ...F MeasurementConsiderations ...
Page 308: ...G GPIB488 1Protocol ...
Page 312: ...H ExamplePrograms ...
Page 317: ...I ContinuousPulseMode ...