Model 2520 User’s Manual
Source-Measure Concepts
5-7
Figure 5-6
Remote staircase sweep mode pulse parameters
Pulse rise and fall times
As shown in
, there are two additional pulse characteristics that require discus-
sion: rise time and fall time. (Characteristics are exaggerated for clarity.) Note that these
characteristics depend on cable connection configuration, setting and range, load, and
operating mode. See the specifications in
Figure 5-7
Pulse characteristics
The rise and fall times of the current source pulse can be controlled with the
:SOURce1:PULSe:TRANsition command, as covered in
. This command allows you to increase the rise time of the pulse. Increasing the rise
time will help reduce overshoot and/or oscillations that could occur when long cables are
used, or if poor connections are present in the test setup.
SOUR1:CURR:START
SOUR1:CURR:LOW
SOUR1:CURR:STEP
SOUR1:CURR:STOP
0A
Returns to 0A
because pulse is
less than Low.
SOUR1:PULS:WIDT
SOUR1:PULS:DEL
Start
Rise
Time
Fall
Time
Current Pulse
(Exaggerated for Clarity)
10%
90%
Summary of Contents for 2520