14-52
SCPI Command Reference
Model 2520 User’s Manual
POSetup
:SYSTem:POSetup <name>
Program power-on defaults
Parameters
<name> =
RST
Power-up to *RST defaults
PRESet
Power-up to :SYSTem:PRESet defaults
SAV0
Power-up to setup stored at memory location 0
SAV1
Power-up to setup stored at memory location 1
SAV2
Power-up to setup stored at memory location 2
SAV3
Power-up to setup stored at memory location 3
SAV4
Power-up to setup stored at memory location 4
Query
:POSetup?
Query power-on setup
Description
This command is used to select the power-on defaults. With RST
selected, the instrument powers up to the *RST default conditions. With
PRES selected, the instrument powers up to the :SYStem:PRESet
default conditions. Default conditions are listed in the SCPI tables
(
).
With the SAV0-4 parameters specified, the instrument powers-on to the
setup that is saved in the specified location using the *SAV command.
Error queue
NOTE
See
for details on the error queue.
[:NEXT]?
:SYSTem:ERRor[:NEXT]?
Read oldest error (code and message)
Description
As error and status messages occur, they are placed in the Error Queue.
The Error Queue is a first-in, first-out (FIFO) register that can hold up to
10 messages. After sending this command and addressing the Model
2520 to talk, the oldest message is sent to the computer and is then
removed from the queue.
NOTE
The :STATus:Queue? command performs the same function as
:SYSTem:ERRor[:NEXT]?. See
“STATus subsystem,” page 14-49
ALL?
:SYSTem:ERRor:ALL?
Read all errors (codes and messages)
Description
This query command is similar to the [:NEXT]? command except that
all messages in the Error Queue are sent to the computer when the
Model 2520 is addressed to talk. All messages are removed from the
queue.
Summary of Contents for 2520
Page 250: ...A Specifications ...
Page 254: ...B StatusandErrorMessages ...
Page 262: ...C DataFlow ...
Page 266: ...D IEEE 488BusOverview ...
Page 281: ...E IEEE 488andSCPI ConformanceInformation ...
Page 285: ...F MeasurementConsiderations ...
Page 308: ...G GPIB488 1Protocol ...
Page 312: ...H ExamplePrograms ...
Page 317: ...I ContinuousPulseMode ...