Model 2520 User’s Manual
SCPI Signal-Oriented Measurement Commands 13-3
NOTE
See
for a detailed explanation on how data flows through the vari-
ous operation blocks of the Model 2520. It clarifies the types of readings that are
acquired by the various commands to read data.
[:SENSe[1]]:DATA[:LATest]?
[:SENSe2]:DATA[:LATest]?
[:SENSe3]:DATA[:LATest]?
Description
These commands work exactly like FETCh?, except that they return
only the most recent reading for the specified SENSe channel.
READ?
Description
This command is used to trigger and acquire readings. The number of
readings depends on how the unit is configured. If the unit is in the fixed
mode, this command will trigger and acquire one set of readings. How-
ever, if the unit is in the sweep mode, this command will trigger the
sweep and then acquire all sweep data points.
When this command is sent, the following commands execute in the
order they are presented:
• :INITiate
• :FETCh?
The :INITiate command starts operation by taking the instrument out of
idle.
NOTE
:READ? is illegal when the :TRIG:SOUR BUS command is in effect. See
,” for details.
After all source and measure operations are completed, the Model 2520
goes back into idle at which time the :FETCh? command is executed.
The readings are sent to the computer and displayed when the Model
2520 is addressed to talk.
NOTE
The :FORM:ELEM command controls the reading source. See
“
,” for details.
MEASure?
Description
This command performs essentially the same function as the :READ?
command, except that it also turns the source outputs on if they were
off.
Summary of Contents for 2520
Page 250: ...A Specifications ...
Page 254: ...B StatusandErrorMessages ...
Page 262: ...C DataFlow ...
Page 266: ...D IEEE 488BusOverview ...
Page 281: ...E IEEE 488andSCPI ConformanceInformation ...
Page 285: ...F MeasurementConsiderations ...
Page 308: ...G GPIB488 1Protocol ...
Page 312: ...H ExamplePrograms ...
Page 317: ...I ContinuousPulseMode ...