Model 2520 User’s Manual
SCPI Command Reference
14-55
RS-232 interface
NOTE
The following commands are intended for use over the RS-232 interface, but
they can also be used over the GPIB.
LOCal
:SYSTem:LOCal
Take Model 2520 out of remote
Description
Normally, during RS-232 communications, front panel keys are opera-
tional. However, the user may wish to lock out front panel keys during
RS-232 communications. See “
This action command is used to remove the Model 2520 from the
remote state and enables the operation of front panel keys in a manner
similar to the GPIB GTL command.
REMote
:SYSTem:REMote
Place the Model 2520 in remote
Description
This action command is used to place the Model 2520 in the remote
state in a manner similar to the GPIB REN command. In remote, the
front panel keys will be locked out if local lockout is asserted. See
“
RWLock
:SYSTem:RWLock
Disable or enable front panel keys
Description
This command is used to enable local lockout in a manner similar to the
GPIB LLO command (See
). When enabled, the front panel
keys except OUTPUT OFF are locked out (not operational) when the
instrument is in remote. (See “
”.) When disabled, the front
panel LOCAL key is operational in remote.
Removing the instrument from remote (:SYST:LOCal) restores front
panel keys operation.
Reset timestamp
RESet
:SYSTem:TIME:RESet Reset
timestamp
Description
This action command is used to reset the absolute timestamp to 0 sec-
onds. The timestamp also resets to 0 seconds every time the Model 2520
is turned on.
Summary of Contents for 2520
Page 250: ...A Specifications ...
Page 254: ...B StatusandErrorMessages ...
Page 262: ...C DataFlow ...
Page 266: ...D IEEE 488BusOverview ...
Page 281: ...E IEEE 488andSCPI ConformanceInformation ...
Page 285: ...F MeasurementConsiderations ...
Page 308: ...G GPIB488 1Protocol ...
Page 312: ...H ExamplePrograms ...
Page 317: ...I ContinuousPulseMode ...