Model 2520 User’s Manual
SCPI Command Reference
14-51
ENABle <list>
:STATus:QUEue:ENABle <list>
Enable messages for Error Queue
Parameters
<list> = (numlist)
where numlist is a specified list of messages that you wish to enable for
the Error.
Query
:ENABle?
Query list of enabled messages
Description
On power-up, all error messages are enabled and will go into the Error
Queue as they occur. Status messages are not enabled and will not go
into the queue. This command is used to specify which messages you
want enabled. Messages not specified will be disabled and prevented
from entering the queue.
DISable <list>
:STATus:QUEue:DISable <list>
Disable messages for Error Queue
Parameters
<list> = (numlist)
where numlist is a specified list of messages that you wish to disable for
the Error Queue.
Query
:DISable?
Query list of disabled messages
Description
On power-up, all error messages are enabled and will go into the Error
Queue as they occur. Status messages are not enabled and will not go
into the queue. This command is used to specify which messages you
want disabled. Disabled messages are prevented from going into the
Error Queue.
SYSTem subsystem
The SYSTem subsystem contains miscellaneous commands that are summarized in
Default conditions
PRESet
:SYSTem:PRESet
Return to :SYSTem:PRESet defaults
Description
This command returns the instrument to states optimized for front panel
operation. :SYSTem:PRESet defaults are listed in the SCPI tables
(
).
Summary of Contents for 2520
Page 250: ...A Specifications ...
Page 254: ...B StatusandErrorMessages ...
Page 262: ...C DataFlow ...
Page 266: ...D IEEE 488BusOverview ...
Page 281: ...E IEEE 488andSCPI ConformanceInformation ...
Page 285: ...F MeasurementConsiderations ...
Page 308: ...G GPIB488 1Protocol ...
Page 312: ...H ExamplePrograms ...
Page 317: ...I ContinuousPulseMode ...