Model 2520 User’s Manual
SCPI Command Reference
14-29
Description
These commands are used to select the current measurement range for
detector #1 and detector #2. The range is selected by specifying the
expected reading. The instrument will then go to the most sensitive
reading that will accommodate that reading. For example, if you expect
a reading of approximately 60mA, then simply let <n> = 0.06 (or 6e-2)
in order to select the 100mA range.
You can also use the UP and DOWN parameters to select range. Each
time UP or DOWN is sent, the next higher or lower measurement range
is selected. When on the maximum range, sending UP is a No-Op (no
operation). When on the lowest range, sending DOWN is a NO-Op.
Select polarity
POLarity <name>
[:SENSe[1]]:VOLTage:POLarity <name>
Set laser diode measurement polarity
:SENSe2:CURRent:POLarity <name>
Set detector #1 measurement polarity
:SENSe3:CURRent:POLarity <name>
Set detector #2 measurement polarity
Parameters
<name> =
POSitive
Positive polarity
NEGative
Negative polarity
Query
:POLarity?
Query measurement polarity
Description
These commands are used to select the measurement polarity for laser
diode voltage measurements and detector current measurements.
POSitive selects normal polarity, while NEGative selects reversed polar-
ity.
See
, “
polarity.
Query voltage limit
TRIPped?
[:SENSe[1]]:VOLTage:PROT:TRIPped?
Query if voltage limit reached
Query
:TRIPped?
Query if voltage limit reached
Description
This query command is used to determine if the laser diode voltage
measurement compliance limit has been reached. A returned value of 1
indicates that the voltage limit has been reached (see
“PROTection[:LEVel] <NRf>,” page 14-35
). A returned value of 0 indi-
cates that the limit has not been reached.
Summary of Contents for 2520
Page 250: ...A Specifications ...
Page 254: ...B StatusandErrorMessages ...
Page 262: ...C DataFlow ...
Page 266: ...D IEEE 488BusOverview ...
Page 281: ...E IEEE 488andSCPI ConformanceInformation ...
Page 285: ...F MeasurementConsiderations ...
Page 308: ...G GPIB488 1Protocol ...
Page 312: ...H ExamplePrograms ...
Page 317: ...I ContinuousPulseMode ...