14-58
SCPI Command Reference
Model 2520 User’s Manual
Abort source/measure cycle
ABORt
Abort operation
Description
When this action command is sent, the Model 2520 aborts operation and
returns to the idle state.
A faster way to return to idle is to use the DCL or SDC command.
Program trigger model
COUNt <n>
:TRIGger[:SEQuence[1]]:COUNt <n>
Set trigger count
Parameters
<n> =
1 to 5000
Specify count
DEFault
Sets count to 1
MINimum
Sets count to 1
MAXimum
5000
INFinite
Run continuously
NOTE
The product of the trigger count and the number of sweep points
(SOUR1:SWE:POIN) cannot exceed 5000.
Query
:COUNt?
Queries programmed count
:COUNt? DEFault
Queries *RST default count
:COUNt? MINimum
Queries lowest allowable count
:COUNt? MAXimum
Queries largest allowable count
Description
This command is used to specify how many times the unit cycles
through the trigger layer. For example, if the trigger count is set to 2, the
unit will cycle through the trigger layer twice.
TRIGger:COUNt INFinite can be used for repetitive source waveforms
or for long tests where only the last reading is important.
SOURce <name>
:TRIGger[:SEQuence[1]]:SOURce <name>
Specify trigger event control source
Parameters
<name> =
IMMediate
Pass operation through immediately
TLINk
Select Trigger Link trigger as event
TIMer
Select timer as event
MANual
Select manual event
BUS
Select bus trigger as event
NSTest
Select low SOT pulse as event
PSTest
Select high SOT pulse as event
BSTest
Select high or low SOT pulse as event
Summary of Contents for 2520
Page 250: ...A Specifications ...
Page 254: ...B StatusandErrorMessages ...
Page 262: ...C DataFlow ...
Page 266: ...D IEEE 488BusOverview ...
Page 281: ...E IEEE 488andSCPI ConformanceInformation ...
Page 285: ...F MeasurementConsiderations ...
Page 308: ...G GPIB488 1Protocol ...
Page 312: ...H ExamplePrograms ...
Page 317: ...I ContinuousPulseMode ...