8 Part replacement instructions
3.
Disconnect the pivot movement sensor.
4.
Remove the detector of the pivot movement sensor.
5.
Remove the pivot movement sensor.
6.
Clean the place of the linear movement sensor with a suitable detergent.
NOTICE!
No glue residue or grease must be left on the surface where the
linear movement sensor is attached.
7.
Attach the new linear movement sensor in place with glue.
NOTICE!
Make sure that the sensor has been evenly attached to the suface.
8.
Connect the linear movement sensor.
9.
Put the new pivot movement sensor in place.
10.
Attach the detector of the pivot movement sensor.
11.
Re-run the motion calibrations. Refer to chapters Device movement calibration and
Z-movement calibration in User and installation manual.