7 Troubleshooting
Error code
Problem
Solution
E18202
Wrong state
Software error. Take logs and boot the
device.
E18204
Drive aborted timeout
Software error. Take logs and boot the
device.
E18205
Opto not found
1. Check connections:
* J4002, J1011, J1039
If fail (open) => replace emergency
stop with cable
* J1011, J1039
If fail (open) => replace cable column
cable extension
2. Replace opto fork
3. Replace i4000.
E18206
No pulses
1. Check connections:
* J1037, J4002
If fail => Replace column cable
extension
2. Replace i4000
3. Replace column.
E18402
Overheat
Let the device cool down.
E18404
Proximity disabled or no pulses
Software error. Take logs and boot the
device.
E18405
Drive aborted button
Software error. Take logs and boot the
device.
E19002
Self-calibration failed
1. Check that there are no objects on
the UI panel.
If fail => remove object
2. Replace the upper cover of the
lower shelf (contains the UI panel
module).
E19003
Invalid state shift
Software error. Take logs and boot the
device.
E19004
Self-diagnostic failed
1. Check connections:
* J4007, J6001
If fail => replace user interface cable
2. Replace the upper cover of the
lower shelf (contains the UI panel
module)
3. Replace i4000.
80
ORTHOPANTOMOGRAPH
™
OP 3D