7 Troubleshooting
Error code
Problem
Solution
E26203
Unexpected opto hit
1. Actuator pushed externally.
2. See additional errors:
* E09011 Opto missing
* E09017 Resistive sensor missing.
E26204
Missed opto
1. Motion blocked externally.
2. See additional errors:
* E09011 Opto missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
E26205
Calibration missing edges
1. Motion blocked externally.
2. See additional errors:
* E09011 Opto missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
E26206
Calibration extra edges
1. Actuator pushed externally.
2. See additional errors:
* E09011 Opto missing
* E09017 Resistive sensor missing.
E26207
Calibration invalid edge
Software error. Take logs and boot the
device.
E26402
Position guard failed
Unintended angle change of the upper
shelf.
E26403
Timeout
Software error. Take logs and boot the
device.
E27202
Position fail
1. Pushed/blocked externally.
2. See additional errors:
* E09012 Opto missing
* E09018 Resistive sensor missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
E27203
Unexpected opto hit
1. Actuator pushed externally.
2. See additional errors:
* E09012 Opto missing