7 Troubleshooting
Error code
Problem
Solution
E54203
Event timeout
Software error. Take logs and boot the
device.
E54204
Button up during exposure
Software error. Take logs and boot the
device.
E54205
Run cannot lock settings
Software error. Take logs and boot the
device.
E54206
Z calibration without homing
Software error. Take logs and boot the
device.
E54402
Button up during start
Software error. Take logs and boot the
device.
E54403
Unhandled event
Software error. Take logs and boot the
device.
E55002
Workspace locked
Software error. Take logs and boot the
device.
E55003
Invalid state
Software error. Take logs and boot the
device.
E55004
Locked
Software error. Take logs and boot the
device.
E56002
Invalid channel
Software error. Take logs and boot the
device.
E56003
Timer init
Software error. Take logs and boot the
device.
E56004
Timer start
Software error. Take logs and boot the
device.
E57002
Failed to send message
Error source i3000:
1. Check connections:
* J3009, J2010
If fail => replace main cable
2. Replace i2000
3. Replace i3000
Error source, i2000 or i3000:
1. Check connections:
* J4009, J3010
If fail => replace twister cable
assembly
* J3009, J2010
If fail => replace main cable
2. Replace i3000
90
ORTHOPANTOMOGRAPH
™
OP 3D