4 Device description
Refer to User and Installation manual, chapter Mains power configuration for detailed
instructions.
NOTICE!
In i4100 back panel board there is a switch (SW1) that is used for
changing, for example, from permanent to plug connection. Refer to User and
Installation manua, chapter Mains power configuration.
4.3.6 Fuses
The allowed main fuse types are listed in the table below. See also
Fusing diagram
.
Table 1: Main fuses
Location
Ref
Value
Type
i4100
220-240Vac
(6.3x32mm)
Littelfuse 215 (Time-Lag) 10A
Cooper Bussman (Time Delay)
S505H-10-R
Back panel
F1 & F2
100-120Vac
(6.3x32mm)
Littelfuse 326 (SlowBlow) 15A
Cooper Bussman (Time Delay)
MDA-15
External warning light
fuse (i4100)
F3
Cooper Bussmann (Time Delay)
S506-2-R
For fuses in circuit boards, see:
i2000 image capture board
i3000 Main control board
i4000 power supply board
ORTHOPANTOMOGRAPH
™
OP 3D
21