7 Troubleshooting
Error code
Problem
Solution
E09008
UI panel missing
1. Check connections:
* J4007, J6001
If fail (open) => replace user
interface cable
2. Replace i6000
3. Replace i4000.
E09009
Z motor opto missing
1. Check connections:
* J4002, J1011, J1039
If fail (open) => replace emergency
stop with cable
* J1011, J1039
If fail (open) => replace column cable
extension
2. Replace opto fork
3. Replace i4000.
E09010
Back panel board missing
1. Check connections:
* J4013, J4101
If fail (open) => replace back panel
cable
2. Replace i4100
3. Replace i4000.
E09011
Pivot opto missing
1. Check connections:
* J3005, J1026
If fail (open) => replace pivot sensor
movement cable
2. Replace opto fork
3. Replace i3000.
E09012
Linear opto missing
1. Check connections:
* J3004, J1025
If fail (open) => replace linear sensor
cable
2. Replace opto fork
3. Replace i3000.
E09013
Rotator opto missing
1. Check connections:
* J2001, J1027
If fail (open) => replace linear sensor
cable
2. Replace opto fork
3. Replace i2000.
66
ORTHOPANTOMOGRAPH
™
OP 3D