6 Maintenance
If you are using the Display Calibration Kit purchaceable from PaloDEx Group Oy, refer to
the Display Calibration Kit Instructions.
6.2 Calibrations
6.2.1 When to calibrate the device
The device must be calibrated and, if necessary, adjusted at regular intervals in
accordance with the national regulations regarding the use, maintenance and service of
dental X-ray devices.
The device must be re-calibrated completely at least
once a year
, by a service
technician according to ORTHOPANTOMOGRAPH
™
OP 3D User and Installation manual.
The last calibration date for each program is listed in the calibration menu.
The device must be calibrated after the first installation or when parts are replaced:
QC images have to be taken always after calibration.
NOTICE!
Refer to the OP 3D User and installation manual for the complete
calibration instructions.