8 Part replacement instructions
1.
Detach the connectors.
2.
Remove the screws (seven pcs).
3.
Lift the board off.
4.
Place the new i2000 board onto its place.
5.
Attach the screws (seven pcs) and tighten them.
6.
Attach the cables and connectors.
7.
For device calibrations, see
When to calibrate the device
.
8.3.8 Replacing the i3000 main control board
NOTICE!
You must remove the upper shelf right and left cover before this
procedure.
Required tools: screwdriver
1.
Remove the connectors from both ends of the i3000 main control board.
ORTHOPANTOMOGRAPH
™
OP 3D