Note
The motor can only operate with retention ring ①
Mat. no. 1.001.5029
.
1
Note
Only the light function is activated and a 1:1 hand and angle piece is displayed with
hand and angle pieces with a metal snap-in button. The torque is not displayed.
5.5.3
Request maximum torque (free use)
▶ Hold down the enter button after the motor stops.
The maximum torque ① of the last motor operation is displayed in Ncm.
Note
The torques are overwritten when the motor restarts under a minimum load of 5
Ncm.
5.5.4
Request maximum torque (program)
See also: 5.4.8 Request maximum torque, Page 39
5.5.5
Recommended programming when setting a number of im‐
plants one after the other
Mode of operation:
Perform all tasks for drilling an implant in Free use mode. Screw in the respective
implant in PRG steps (max. 6). At the end of the task, the torque screw-in value can
be requested and manually documented for evaluating primary stability (in the form
of the following table, for example).
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.5 Functions exclusively with the 300 plus program version
43/66