▶ Start the motor. The motor always rotates to the right.
Note
Do not apply any pressure.
The motor may not experience any additional load.
The process can be monitored on the display.
After calibration, the following appears on the display:
If calibration is unsuccessful, the following appears on the display:
Note
The output values are retained in case of interruption or error.
The calibration automatically ends. The INTRAsurg 300 plus is ready with the new
setting.