▶ Connect the new and existing coolant hose ① to the coupling ③.
▶ Carefully remove the old hose from the motor end and simultaneously shove on
the new one until the coupling ③ piece is visible.
▶ Pull the coupling ③ out of the hose opening e.g. using a tweezers so that the
coupling piece is not clamped there.
7.2.3
Disposable coolant hose
Sterile-packaged coolant hoses can be ordered under Mat. no. 1.001.9902 (pack‐
age contents: 10 units).
This should be used for applications with an increases infection risk or when the
motor hose cannot be prepared according to the existing instructions.
The disposable coolant hose consists of an insertion tip, pump hose, clip and hose
holder.
The length of the hose can be cut at the exit end with a sterile knife or sterile scissors.
Observe the instructions regarding use on the package.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
7 Troubleshooting | 7.2 Exchange coolant hose
59/66