4.2
Unpacking
Note
Do not kink the motor hose since it may damage it.
Note
Save the box with all the packaging materials so that the device can be safely
shipped (for example for a yearly service check). Always send both hose boxes so
that the device can be safely packaged to prevent damage.
Prevent damage by properly removing from the package.
▶ Open box.
▶ Remove the hose box. Remove the multifunctional foot control and additional
equipment.
Note
When removing, note that the multifunctional foot control is tightly connected to the
device.
▶ Pull out the INTRAsurg 300/300 plus vertically upward and place on a flat sur‐
face.
4.3
Adjust the mains input voltage
CAUTION
Voltage-conducting parts
Electrical shock.
▶ Unplug the power cable from the power supply.
A country-specific mains input voltage (a rating play) is set at the plant. Check the
setting and change it if necessary to the existing mains input voltage.
The voltage selector ① can be set to 100V, 120V or 230V. The existing setting can
be read at the cutout ④.
Remove the drawer ③.
Remove the fuse holder ② from the fuse drawer ③, and turn it so that the required
voltage value appears in the cutout ④.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
4 First use | 4.2 Unpacking
19/66