Step
Instrument
Speed (rpm) Torque
(Ncm)
Pump
Motor r/l
1
27:1
11 – 50
40
0
->
2
27:1
11 – 50
40
0
->
3
27:1
11 – 50
40
0
->
4
27:1
11 – 50
40
0
->
5
27:1
11 – 50
40
0
->
6
27:1
11 – 50
40
0
->
5.5.6
INTRAsurg calibration
The influence of the efficiency of the instruments is taken into account when testing
the torque display.
Store the INTRAsurg calibration device at 15°C to 30°C to provide precise calibra‐
tion. The calibration value refers to the temperature of the INTRAsurg calibration of
23°C. If this temperature is not maintained, the value needs to be corrected ac‐
cording to table. For example: Calibration at 20°C (read 15.2) results in a corrected
input value (+0.3) of 15.5.
Tempera‐
ture
15℃ 16℃ 17℃ 18℃ 19℃ 20℃ 21℃ 22℃ 23℃ 24℃ 25℃ 26℃ 27℃ 28℃ 29℃ 30℃
correc‐
tion
+0.8 +0.7 +0.6 +0.5 +0.4 +0.3 +0.2 +0.1 0
-0.1 -0.2 -0.3 -0.4 -0.5 -0.6 -0.7
A sterile instrument can also be used for calibration. To prevent contamination, use
a sterile catch pin.
Note
The INTRAsurg Calibration system can be disinfected.
▶ Pre-position cleaned and set up angle piece 27:1 and INTRAsurg Calibration
Mat. no. 1.002.3569.
▶ Press the parameter button for 3 seconds.
The following message appears on the display:
▶ Read the calibration affixed to the bottom of the INTRAsurg Calibration (such as
15.2), correct the value according to the temperature, and enter with the Plus
and Minus keys.
▶ Confirm with the enter button.
▶ Insert the catch pin into the INTRAsurg Calibration.
▶ Place the INTRAsurg Calibration on a fixed base and hold.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.5 Functions exclusively with the 300 plus program version