Changed values can be saved in the parameter menu and are then available for the
next use.
The following values can be changed in the parameter menu:
▪ Transfer ratio of the handpiece and contra-angle handpiece
▪ max. speed
▪ Direction of motor rotation
▪ Coolant pump setting
▪ max. torque
See also: 5.3 Free use, Page 28
CAUTION
Incorrect use.
danger of injury.
▶ Always check the values before use.
5.4.5
Limit the number of steps
Six steps are saved upon first start up and after setting up the delivery status.
Example: Limiting the steps
The number of steps should be limited to four.
▶ Select step 4 ①.
PRG
▶ Press the program button for at least three seconds.
An acoustic signal confirms the entry.
Note
Now only the first four steps can be selected.
When you continue to step 4, the message
END
appears.
If you continue, you go to free use.
Remove restriction
▶ Select the last step ① (4 in this example).