Battery Power Line II User’s Manual DePuy Synthes Power Tools 65
Problem
Possible Causes
Solution
Drive unit does not start
No battery in the drive unit
Insert charged battery
Battery is discharged
Charge or replace battery
Battery is defect
Replace battery
Drive unit is defect
Send drive unit to DePuy Synthes
service center
Drive unit did not cool down
after sterilization
Allow to cool to room temperature
Mode switch is set to "lock" (off position)
Set mode switch to ON/FWD/REV
No electrical contact between drive unit
and battery casing
Reinsert or change battery casing
Drive unit lacks power
Battery is not fully charged or past life cycle Charge or replace battery
Wrong attachment used (eg, drilling speed
vs reaming speed)
Change attachment
Drive unit has not been properly serviced
Send drive unit to DePuy Synthes
service center
Attachments have not been properly serviced
Send attachments to DePuy Synthes
service center
Drive unit suddenly stops
Drive unit has overheated
Allow to cool to room temperature
Battery is empty/discharged
Charge or replace battery
Drive unit is defective
Send drive unit to DePuy Synthes
service center
Drive unit continues to run
after releasing trigger
Trigger is jammed by residue
Immediately turn mode switch to
"lock" (off position) or remove the
battery casing
Precaution: Clean and lubricate
trigger according to care and
maintenance guidelines.
Drive unit is defective
Immediately turn mode switch to
"lock" (off position) or remove the
battery casing. Send drive unit
to DePuy Synthes service center
General
Troubleshooting