Battery Power Line II User’s Manual DePuy Synthes Power Tools 67
Problem
Possible Causes
Solution
Attachments cannot couple
to drive unit
Coupling is blocked by residue
Precaution: Immediately turn mode
switch OFF (Lock position). Remove
solid particles with pickups. Clean
and lubricate according to care and
maintenance guidelines.
Attachment coupling is damaged
Send damaged attachment to
DePuy Synthes service center
Difficulty removing
attachments from drive unit
Coupling is blocked by residue
Precaution: Immediately turn mode
switch OFF (Lock position). Remove
solid particles with pickups. Clean
and lubricate coupling sleeve
according to care and maintenance
guidelines.
Drive unit coupling sleeve is damaged
Send damaged drive unit to
DePuy Synthes service center
Bone and drive unit heat up
during surgery
The cutting tool is blunt
Replace the tool
Battery Oscillator II
Problem
Possible Causes
Solution
Saw blade is difficult to
couple or cannot be coupled
General wear and tear has affected the
connection geometry of the saw blade
Replace the saw blade
Bone and drive unit heat up
during surgery
The cutting teeth of the saw blade
are blunt
Replace the saw blade
Battery Oscillator II vibrates
too intensively
Saw blade locking mechanism is not tight
Tighten the locking knob on the saw
blade quick coupling
Battery Reamer/Drill II
Troubleshooting