44 DePuy Synthes Power Tools Battery Power Line II User’s Manual
Care and Maintenance
Automated Cleaning Instructions with
Manual Precleaning
Notes
• Manual precleaning prior to automated cleaning
is important to ensure cannulations and other
difficult to access areas are clean.
• Alternative cleaning procedures other than in
the procedure described below (including manual
precleaning) have not been validated by
DePuy Synthes.
1. Remove Debris
Rinse the device under running cold tap water for a
minimum of 2 minutes. Use a sponge, soft, lint-free cloth
or soft-bristled brush to assist in removing gross soil
(Figure 1). For cannulations of the handpiece and
attachments, the cleaning brush (516.101) should be used.
Notes
• Do not use pointed objects for cleaning.
• Brushes and other cleaning tools shall be either
single-use items or, if reusable, be decontaminated
at least daily using a solution as detailed
in section 3. “Spray and wipe.” Brushes should be
inspected before daily use and discarded if they
have degraded to the point where they may
scratch instrument surfaces or be ineffective due
to worn or missing bristles.
Precaution: Never immerse the handpiece, batteries,
battery casing, or attachments in aqueous solutions
or in an ultrasonic bath or use pressurized water as
this will cause damage to the system.
Figure 1
Cleaning brush (516.101)
(310 mm)
(78 mm)
(55 mm)
(
B
4.7 mm)