Battery Power Line II User’s Manual DePuy Synthes Power Tools 45
2. Manipulate Moving Parts
Manipulate all moving parts such as triggers, sliding
sleeves, attachment release rings, saw blade coupling,
and switches under running tap water to loosen and re-
move gross debris.
3. Spray and Wipe
Spray and wipe the device using a neutral pH enzymatic
solution for a minimum of 2 minutes (Figure 2). Follow
the enzymatic detergent manufacturer’s directions for
correct temperature, water quality (ie, pH, hardness)
and concentration/dilution.
4. Clean with Detergent
Clean the device manually under running warm water
using an enzymatic cleaner or detergent for a minimum
of 5 minutes. Manipulate all moving parts under running
water. Use a soft-bristled brush and/or soft, lint-free cloth
to remove all visible soil and debris (Figures 3 and 4). Follow
the enzymatic cleaner or detergent manufacturer’s
instructions for use for correct temperature, water quality,
and concentration/dilution.
Note: For the Quick Coupling for Pins 3.2 mm
diameter (530.796), the brush should only be
inserted from the front.
5. Rinse with Tap Water
Rinse the device thoroughly using cool to lukewarm
running water for a minimum of 2 minutes. Use a syringe
or pipette to flush lumens and channels. Actuate joints,
handles, and other movable device features to rinse
thoroughly under running water.
6. Visually Inspect Device
Inspect the cannulations, sliding sleeves, attachment
release rings, etc for visible soil. Repeat steps 1–6 if
visible soil remains.
Care and Maintenance
Automated Cleaning Instructions with Manual Precleaning
Figure 2
Figure 3
Figure 4