4. Wrapping and Labeling
The batteries can either be wrapped or peel pouched. Fol-
low the instructions for wrapping and labeling with
the STERRAD
®
Systems provided by Advanced Sterilization
Products or with the V-PRO
®
Systems provided by STERIS.
5. Sterilization
Process the battery in the STERRAD or V-PRO Systems fol-
lowing the instructions provided by the legal manufacturer
Advanced Sterilization Products or STERIS.
The BPL II Li-Ion battery (530.630) has been validated with
the following:
STERRAD Systems
and Cycle
V-PRO Systems
and Cycle
STERRAD 50
STERRAD 50
V-PRO 1
V-PRO 1
Lumen Cycle
Lumen Cycle
STERRAD 200
STERRAD 200
Short Cycle
Short Cycle
V-PRO 1 Plus
V-PRO 1 Plus
Lumen Cycle,
Lumen Cycle,
Non Lumen Cycle
Non Lumen Cycle
STERRAD 100S
STERRAD 100S
Short Cycle
Short Cycle
V-PRO maX
V-PRO maX
Lumen Cycle, Non Lumen Cy-
Lumen Cycle, Non Lumen Cy-
cle, Flexible Cycle
cle, Flexible Cycle
STERRAD NX
STERRAD NX
®
®
Standard cycle
Standard cycle
V-PRO 60
V-PRO 60
Lumen Cycle,
Lumen Cycle,
Non Lumen Cycle
Non Lumen Cycle
STERRAD 100NX
STERRAD 100NX
Standard cycle
Standard cycle
Note: No other sterilization methods are allowed.
Care and Maintenance
STERRAD
®
and V-PRO
®
Systems Sterilization Guide
For Li-Ion Battery (530.630)
1. Preparation Prior to Reprocessing
Before cleaning, remove all instruments and attachments
from the power tool. Remove the battery casing from
the handpiece and then remove the battery from the
battery casing.
Inspect battery for cracks or damage.
• Take care not to spray the contacts or touch both
contacts at the same time with the damp cloth due
to danger of short circuiting. Do not immerse
batteries into water or clean them in a washer/
disinfector.
2. Cleaning and Disinfection
To clean the battery, wipe it off with a clean, soft, and
lint-free cloth dampened with deionized water and dry
prior to processing.
To disinfect the battery, wipe it off with a clean, soft and
lint-free cloth dampened with a minimum of 70% alco-
hol-based disinfectant for thirty (30) seconds. A disinfec-
tant that is VAH listed, EPA registered or locally recog-
nized is recommended. This step has to be repeated two
(2) additional times using a new, clean, soft and lint-free
cloth dampened with a minimum 70% alcohol-based dis-
infectant each time. Follow the instructions provided by
the manufacturer of the disinfectant.
3. Charging
1
Place the battery into the Universal Battery Charger II
2
(05.001.204).
Charge the battery. When the green LED illuminates, the
battery is fully charged.
Upon completion of charging the battery, wipe the bat-
tery with a minimum 70% alcohol-based disinfectant
prior to wrapping and labeling.
1 Universal Battery Charger II with firmware version 14.0 or higher is required to
charge BPL II Li-Ion batteries.
2 For detailed information on the Universal Battery Charger II, please refer to the
Instructions for Use for the Universal Battery Charger II.
3 For detailed information regarding Storage and Use, please refer to the
Instructions for Use for the Battery Power Line II.
4 Data on file at DePuy Synthes.
61 DePuy Synthes Power Tools Battery Power Line II User’s Manual