Battery Power Line II User’s Manual DePuy Synthes Power Tools 13
DePuy Synthes non-sterile batteries and advanced charging
technology optimize intraoperative battery capacity,
maximize battery lifespan, and shorten turnaround time.
One Universal Battery Charger II (05.001.204) for multiple
DePuy Synthes battery-driven systems simplifies the
charging process. Simple aseptic technique preserves the
sterile field when assembling the battery pack.
Instruments
530.630
Battery for Battery Power Line II
530.660
Insertion Shield for Battery Power Line II
530.690
Battery Casing for Battery Power Line II
Assembling and Inserting the Battery Pack
For instructions on Assembling and Inserting a battery
sterilized in Sterrad
®
or V-PRO
®
systems, please refer to
page 61.
Scrubbed Person
Open the lid of the battery casing (Figure 1).
Ensure the lid of the battery casing is fully open (Figure 2).
Ensure that the lid of the battery casing is facing towards
the scrubbed person (Figure 3).
Position the insertion shield securely on top of the battery
casing (Figure 4).
Notes
• The insertion shield helps to guide the battery into
the battery casing and prevents contamination
of the sterile casing by the non-sterile battery.
• Sterilize the insertion shield after each use to
ensure aseptic conditions when inserting the
non-sterile battery into the sterile battery casing.
Precautions
• If the non-sterile battery contacts the outside
of the battery casing, the battery casing must be
cleaned and resterilized before being used in the
operating room.
• Do not insert the non-sterile battery into the
battery casing whilst a handpiece is attached.
Operating Instructions
Battery Pack
(battery casing with inserted battery)
Figure 1
Figure 2
Figure 3
Figure 4