Notes:
• Ensure that the STERRAD or V-PRO sterilized
battery does not come into contact with any
liquids or other substances whilst being inserted.
• Following the introduction of the STERRAD or
V-PRO sterilization methods, the markings on the
BPL II Li-Ion batteries (530.630) have been
changed, including replacing “DO NOT
STERILIZE” with “DO NOT AUTOCLAVE MAX.
60°C / 140°F”. However, all BPL II Li-Ion batteries
(530.630) can be STERRAD or V-PRO sterilized,
regardless of their markings.
7. STERRAD
®
or V-PRO
®
Sterilized Battery Insertion
•
The scrubbed person performs all of the steps according
to Figure 1–8
•
Open the lid of the battery casing (530.690) by
simultaneously pressing both battery casing locks
(Fig. 1)
•
Ensure the lid of the battery casing is fully open (Fig. 2)
and facing towards the scrubbed person (Fig. 3)
•
Insert the sterile battery into the battery casing (Fig. 4)
and press it completely into the battery casing to ensure
a correct seat (Fig. 5)
•
Close the battery casing by simultaneously pressing
both battery casing locks and closing the lid
(Fig. 6 and Fig. 7)
•
Attach the battery pack
After usage, remove the battery casing from the
handpiece and open the lid. The circulating person then
removes the battery. Ensure that the battery does not
touch the exterior of the battery casing to avoid
contaminating the battery. Put the battery immediately in
the charger.
STERRAD
®
is a registered trademark of Advanced Sterilization Products (ASP).
V-PRO
®
is a registered trademark of STERIS Corporation.
For full user instructions and precautions please consult the product IFU.
Please contact your DePuy Synthes sales representative for more information.
All surgical techniques are available as PDF files at DSHuddle.com
62 DePuy Synthes Power Tools Battery Power Line II User’s Manual
Figure 1
Figure 3
Figure 5
Figure 7
Figure 2
Figure 4
Figure 6
Figure 8
Care and Maintenance
STERRAD
®
and V-PRO
®
Systems Sterilization Guide