Battery Power Line II User’s Manual DePuy Synthes Power Tools 75
Table 2: Immunity (all devices)
Guidance and manufacturer’s declaration—
electromagnetic immunity
The BPL or BPL II System is intended for use in the
electromagnetic environment specified below. The
customer or the user of BPL or BPL II System should
ensure that it is used in such an environment.
Immunity Test
Standard
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment—Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete, or ceramic tile.
If floors are covered with synthetic material, the relative
humidity should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
n/a
Mains power quality should be that of a typical
commercial or hospital environment
Surge
IEC 61000-4-5
±1 kV line to line
±2 kV line to earth
n/a
Mains power quality should be that of a typical
commercial or hospital environment
Voltage dips, short
interruptions, and
voltage variations on
power supply lines
IEC 61000-4-11
<5 % UT
(0.5 cycle)
40% UT
(5 cycles)
70% UT
(25 cycles)
<5 % UT for 5 s
n/a
Mains power quality should be that of a typical
commercial or hospital environment
Note: UT is the a.c. mains voltage prior to application of the test level.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
100 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment
Electromagnetic Compatibility
Accompanying Documents According to IEC 60601-1-2, 2007, Clause 6