4 DePuy Synthes Power Tools Battery Power Line II User’s Manual
Safety Instructions
The Battery Power Line II (BPL II) is only to be used
for patient treatment after careful consultation of the
instructions for use. It is recommended that an alternative
system is available to use during application, as technical
problems can never be completely ruled out.
The Battery Power Line II is designed for use only by
physicians and trained medical personnel.
DO NOT
use any component if damage is apparent.
DO NOT
use any component if the packaging is damaged.
DO NOT
use this equipment in the presence of oxygen,
nitrous oxide, or a mixture consisting of flammable
anesthetic and air.
To ensure proper operation of the tool, only use
DePuy Synthes original accessories.
Before first and every use and prior to returning for service,
power tools and their accessories/attachments, excluding
the battery, have to be run through the complete
reprocessing procedure. Protective covers and films must
be fully removed before sterilization.
For the tool to function properly, DePuy Synthes
recommends that it is cleaned and serviced after each
use in accordance with the process recommended in the
“Care and Maintenance” section. Compliance with these
specifications can considerably extend the service life
of the tool. Only use Synthes Special Oil (519.97) to
lubricate the tool.
We recommend using new DePuy Synthes cutting tools
for every surgical procedure. Efficiently working cutting
tools are the basis for successful surgery. Therefore, check
used cutting tools after every use for wear and/or damage
and replace them if necessary.
Cutting tools must be cooled with irrigation fluid to
prevent heat necrosis.
The user of the product is responsible for proper use
of the equipment during surgery.
Check proper operation of the tool before using it on
the patient.
To prevent overheating, always respect the specified duty
cycles on page 71.
If the Battery Power Line II is used in conjunction with an
implant system, make sure to consult the corresponding
Technique Guide.
For important information regarding electromagnetic
compatibility (EMC) please refer to the “Electromagnetic
Compatibility” section in this manual.
The tool is classified as type BF against electrical shock and
leakage current. The tool is suitable for use on patients in
accordance with IEC 60601-1.
Note: To ensure proper operation of the tool, annual
maintenance by a DePuy Synthes service center is
necessary. The manufacturer shall assume no
responsibility for damage resulting from improper
operation, and neglected or unauthorized
maintenance of the tool.
Introduction
General Information