Troubleshooting
Problem
Possible Causes
Solution
Cutting tool is difficult to
couple or cannot be coupled
to an attachment
The attachment or cutting tool is deformed
from wear
Replace the attachment or cutting tool, or
send it to a DePuy Synthes service center
Attachment becomes
excessively hot
Attachment is used too long
Allow attachment to cool (see Duty
cycles on page 69)
Rotating attachment turns
too slowly
Wrong attachment used (eg, reaming
speed vs drilling speed)
Change attachment
Kirschner wire cannot be
inserted into the Kirschner
wire attachment
Kirschner wire attachment is not opened
far enough
Fully open the adjustment sleeve on the
attachment, insert the Kirschner wire
and close the adjustment sleeve
Kirschner wire cannot be
secured despite pulling the
tension lever
Kirschner wire attachment is opened
too far
Close the adjustment sleeve on the
attachment until the wire is fixed
Kirschner wire is jammed in
the attachment and cannot
be moved
Kirschner wire was inserted at an angle
and is jammed in the attachment
Send Kirschner wire attachment to
DePuy Synthes service center
Guide pin cannot be inserted
into the front of the Quick
Coupling for Pins attachment
(530.796) or cannot
be grasped
Diameter or shaft geometry is unsuitable
Quick Coupling for Pins (530.796)
allows insertion and removal of 3.2 mm
diameter guide pins with round,
triangular, or flat shafts only
Bone and cutting tool
becomes excessively hot
Cutting tool is blunt
Replace cutting tool
Attachments and Cutting Tools (continued)
If the recommended solutions do not work, send the power
tool to your local DePuy Synthes service center.
For further technical questions or information on our
services, please contact your DePuy Synthes representative.
For troubleshooting for the Universal Battery Charger II,
please consult the relevant instructions for use (J8895).
Troubleshooting