SECTION 10—COMPRESSOR REPLACEMENT
61
8. Disconnect the two wires from the high pressure switch (FIGURE 10.2).
FIGURE 10.2
Compressor Replacement - HomeFill Units Manufactured Before 8/1/05
9. Using the
7
/
16
open ended wrench, loosen the nut and disconnect the high pressure
oxygen line from the cylinder fillport connector (FIGURE 10.3).
10. Loosen the two screws that secure the bottom of the control panel in place
11. Remove the two screws that secure the control panel gussets to the rear of the base
FIGURE 10.3
Compressor Replacement - HomeFill Units Manufactured Before 8/1/05
CAUTION
DO NOT cut the tubing or damage (score) the tubing when removing tie-wraps.
Otherwise, HomeFill unit will not fill cylinders correctly.
Wires
High Pressure Switch
High Pressure Oxygen Line
Cylinder Fillport
Connection
Nut (Underneath)