SECTION 8—BASE REPLACEMENT
44
SECTION 8—BASE REPLACEMENT
Tools Required:
7
/
16
open ended wrench,
5
/
8
open ended wrench, Diagonal Cutters,
Needle Nose Pliers, Phillips screwdriver.
Disassembly
NOTE: For this procedure, refer to FIGURE 8.1 through FIGURE 8.7.
1. Install an empty cylinder onto the compressor to release pressure in lines. Refer to
Connecting/Disconnecting Cylinder to/from Compressor on page 20.
2. Remove cylinder. Refer to Connecting/Disconnecting Cylinder to/from Compressor
3. Disconnect power from the HomeFill unit.
4. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 25.
5. Disconnect the two pin connector that connects the motor to the wiring harness
6. Using needle nose pliers, disconnect the two wires from the capacitor (FIGURE 8.1).
7. Using needle nose pliers, disconnect the grounding clip from the motor (FIGURE 8.1).
FIGURE 8.1
Base Replacement
Two Pin
Connector
(STEP 4)
Wires (STEP 5)
Capacitor
Motor
Grounding
Clip
Control Panel
Gusset
Mounting
Screw
Control Panel Gusset
Mounting Screw
(Hidden behind wiring assembly)