SECTION 7—BASE COMPONENTS
41
Removing/Installing Connector Fillport Dust Cover and
Lanyard - Units Manufactured After 10/02/2005
NOTE: For this procedure, refer to FIGURE 7.8 on page 41.
Tools required for this procedure are:
•
7
/
16
Wrench
• Phillips screwdriver
•
9
/
64
Allen Wrench
Removing
1. Remove the locknut connecting the oxygen line to the connector fillport.
2. Remove the mounting screw that connects the bracket to the base assembly.
3. Remove the mounting screw that connects the flange to the base assembly.
4. Remove the quick-release pin and remove the connector fillport and two washers
from the compressor.
5. Remove the three mounting screws at the bottom of the connector fillport.
6. Remove the flange.
7. Remove the existing lanyard from the shaft.
Installing
1. Install the new lanyard onto the shaft.
2. Secure the flange to the bottom of the connector fillport with the three mounting
screws. Securely tighten.
3. Secure the two washers and connector fillport to the compressor with the quick release
pin.
4. Secure the flange to the base assembly with a mounting screw. Securely tighten.
5. Secure the bracket to the base assembly with a mounting screw. Securely tighten.
6. Connect the oxygen line to the connector fillport with the locknut. Securely tighten.
FIGURE 7.8
- Removing/Installing Connector Fillport Dust Cover and Lanyard - Units
Flange
Mounting Screws
Bracket
Mounting Screw
Mounting Screw
Shaft
NOTE: Lanyard not shown.
Locknut for Oxygen Line
Connector Fillport
Quick Release Pin
Washers
Dust Cover