SECTION 13—TROUBLESHOOTING/CYLINDER FILL TIMES
89
FIGURE 13.1
Timer
WARNING
Use extreme caution when testing this device. It may start unexpectedly, and it
contains many pinch points.
DO NOT spray electrical connections with leak-tek.
NEVER allow HomeFill to continue running above 2100 psi (14479 kPa).
Before testing begins, make sure the test compressor is in good condition.
Perform leak checks on the HomeFill compressor throughout testing.
Perform periodic maintenance on test concentrator to ensure consistent results.
1. Place an empty M6 cylinder into the connector fillport on the HomeFill. Refer to
Connecting the Cylinder to the Compressor on page 20.
2. Ensure that the flow selector on cylinder is positioned at “0”.
3. Connect the concentrator bulkhead fitting to the HomeFill inlet fitting. Refer to
Concentrator Verification and Maintenance on page 16.
4. Attach the timer to the HomeFill:
A. Note the orientation of the BROWN and BLUE HomeFill harness wires connected
to the hour meter for later reconnection.
B. Disconnect both HomeFill harness wires from the hour meter.
C. Attach the BROWN HomeFill harness wire to the BROWN timer harness
connector (i.e., BROWN wire to BROWN wire).
D. Attach the BLUE HomeFill harness wire to the BLUE timer harness connector (i.e.,
BLUE wire to BLUE wire).
E. Attach the BROWN timer harness wire to the connector on the hour meter to
which the BROWN HomeFill harness wire is normally connected.
F. Attach the BLUE timer harness wire to the connector on the hour meter to which
the BLUE HomeFill harness wire is normally connected.
5. Plug the HomeFill into 220 VAC power source.
6. Turn the power switch On.