SECTION 7—BASE COMPONENTS
34
SECTION 7—BASE COMPONENTS
Replacing the Fuse
NOTE: For this procedure, refer to FIGURE 7.1.
1. Unplug the power cord from the electrical outlet.
2. Remove the power cord from the power inlet A on the compressor base C.
3. Pull out the fuse drawer
B
at the bottom of the power inlet.
4. Using a fuse puller or a small flat blade screw driver, remove the front fuse inside the
fuse drawer and discard.
NOTE: The fuse drawer contains space for two fuses. The front location is for the operational fuse.
The unit is supplied with the front fuse only. The rear location is for an optional spare fuse. When
replacing the front fuse, ensure that there is a spare fuse with the same type and rating in the fuse
drawer.
5. Install replacement fuse in front fuse
D
slot.
NOTE: Use a 5 mm x 20 mm, time lag, low breaking capacity, 250V, 2 Amp, IEC 60127 style only
(Invacare part number 1099820).
6. Push fuse drawer back into the power inlet.
7. Install power cord into the power inlet on the compressor base.
8. Plug power cord into electrical outlet.
9. Turn the power switch to the On (
|
) position and check to see that the unit powers up
properly.
10. If the unit does not come on, turn the power switch off, unplug the power cord, and
call your dealer, provider or Invacare for service.
FIGURE 7.1