SECTION 8—BASE REPLACEMENT
48
23. Remove the six mounting screws that secure the pump/motor assembly to the base
FIGURE 8.5
Base Replacement
24. Cut the tie wrap that secures the capacitor in place and remove capacitor
NOTE: Located underneath the capacitor is a tie wrap mount.
25. Remove the mounting screw that secures the tie wrap mount (not shown) to the base.
26. Remove the tie wrap mount (not shown).
27. Remove the two mounting screws that secure the transformer to the base
28. Remove the transformer from the base (FIGURE 8.6).
29. Remove the two mounting screws that secure the power inlet to the base
30. Remove the power inlet from the base (FIGURE 8.6).
TOP VIEW WITH CONTROL PANEL ASSEMBLY REMOVED
Mounting Screws
Mounting
Screws
Mounting
Screws
Motor/Pump Assembly
Base