SECTION 11—TIMING BELT REPLACEMENT
77
14. Using diagonal cutters, cut the tie wrap that secures the oxygen inlet hose to the
receptacle fitting (FIGURE 11.4).
15. Remove the oxygen inlet hose from the receptacle fitting (FIGURE 11.4).
16. Using diagonal cutters, cut the tie wrap that secures the oxygen sensor tubing to the
oxygen inlet hose (FIGURE 11.4).
17. Remove the oxygen sensor tubing from the oxygen inlet hose (FIGURE 11.4).
18. Note where each of the four wires and two jumpers are connected to the transformer.
19. Using needle nose pliers, disconnect the connectors from the transformer. See Detail
20. Using needle nose pliers, disconnect the two connectors from the power inlet
21. Using needle nose pliers, disconnect the two connectors (not shown) from the hour
22. Lift control panel assembly off of base and set aside (FIGURE 11.4).